High Flow Calibration
PROCEDURE
(For flow meters providing a voltage or current signal to the X96S.
The flow must be running at the highest speed during the High Calibration.
Using the Ronan local display or HART communicator find the Main Menu;
Scroll down until you reach and highlight Calibration.
Access Calibration by pressing the right arrow key.
Scroll down until you reach and highlight Cal Speed.
Access the Cal Flow Rate by pressing the right arrow key.
Scroll down until you reach and highlight High Calibrate.
Access the High Calibrate by pressing the right arrow key.
Access the High Flow Rate Cal by pressing the right arrow key.
Press the F4 key (OK) to the prompt ‘Warning – Remove loop from automatic control
before doing the calibrate procedure’.
Wait until the prompt changes from
Performing Calibrate procedure’ … (You will also see a timer counting down)
Once the timer reaches zero you will see ‘High Calibrate Flowrate Counts = ‘value”
Press OK to accept or Abort to Cancel calibration.’
Press the F4 key (OK).
Scroll down until you reach and highlight High Flow Rate 5000.000 gal/min. . (Note 5000.000 gal/min
value and units may very depending on the what value and units were selected.) (Note: You must access
the High Flow Rate and either change or acknowledge the value to complete the calibration.)
Access High Flow Rate 5000.0000 gal/min by pressing the right arrow key.
Use the arrow keys to select and change the display value to the correct speed. Once the displayed value is
correct, press the F4 key (Enter).
The Flow Calibration is complete.
For an illustrated procedure, see the following pages.