52
Quick Start Reference – Calibrating Density
(For Completion of a Single Point Reference)
C
F1 F2 F3 F4
Ronan Engineering
X96S Mass Flow Gauge
Main Menu
Variables
Status Display
Configuration
Digital Outputs
Digital Inputs
→
Calibration .
Dia
nostics
Step 3 ‘Low Reference Density’ Calibration
(Have the process ready filled with water)
Scroll down ↓ to “Calibration” Press → to enter menu
Scroll down ↓ to “Cal Density” Press the → to enter menu
Scroll down ↓ to “Low Reference” Press the → to enter menu
Select “Perform Reference” press → to enter menu
Read the next screen carefully and acknowledge by pressing the F4
(OK) Key. Referencing will begin. (This may take several minutes.)
Scroll down ↓ to “Ref Density” Press → to enter menu
Notice the value that is displayed.
Use ← Key and → Key to toggle back and forth between digits.
Change the value using ↑ or ↓
Enter the value of the process (Example 0.0000 %Solids for water).
Press The F4 Key to enter and store the value.
Press the F3 Key (Home) to return to the Main Menu.
C
F1 F2 F3 F4
Ronan Engineering
X96S Mass Flow
Gauge
Low Reference
→
Perform
Reference
.
Ref Density 0.0000
Ref Cap 20000
HOME
C
F1 F2 F3 F4
Ronan Engineering
X96S Mass Flow Gauge
Calibration
State Fully Calibra
Ref Constants
→
Cal
Density .
Cal Flow Rate
Temp Reference
Cal Remote Solids
Last Ref.Date
C
F1 F2 F3 F4
Ronan Engineering
X96S Mass Flow Gauge
Cal Density
State
→
Low Reference
High Calibrate
Manual Entry
Clear Ref/Cal
C
F1 F2 F3 F4
Ronan Engineering
X96S Mass Flow Gauge
Low Reference
WARNING – Remove
loop from automatic
control, verify that
source is on, and the
vessel/pipe is empty.
ABORT OK
C
F1 F2 F3 F4
Ronan Engineering
X96S Mass Flow Gauge
Low Reference
Reference
→
R
ef
Density 0.0000
Ref Cap 18754
HOME
C
F1 F2 F3 F4
Ronan Engineering
X96S Mass Flow Gauge
Ref Density
0.0000 SGU
0.0000
ESC ENTER