X96-2008PL
X96-2008PL is the Digital Input/Output Module. A total of 16 bits of digital I/O and wetting/encoder power is provided by the
module.
8 isolated digital inputs are provided. These inputs can be configured for use as:
dry
6
or live
7
contact monitoring,
quadrature encoder
8
, or
pulse counter.
4 relay (2 Amp capacity) output points are provided. Form “C” outputs are brought out to the connector (three connections per
relay).
4 isolated open collector output points are provided. These outputs are capable of switching 4.5 to 30 Volts (externally supplied)
at a maximum of 50 ma.
24 volts DC is provided to be used as a wetting voltage when needed.
An isolated 15 volt DC power supply capable of providing 200 mA is also provided. The primary use of this power supply is to
power a quadrature encoder but can be used for other purposes if it is not required for this purpose.
X96-2009PL
X96-2009PL is the Scintiallation Detector Interface Module. This optional
9
module provides:
1 isolated scintillation input (pulse counter, max signal 0-12
10
V, threshold 0.6 V)
1 head temperature input (1 uA per deg K)
1 non-isolated RTD (3-wire) input
isolated power for the scintillation detector 24 V 40 mA supply
11
.
6
When used with dry contacts, jumpers shall be used on the connector block to provide the wetting voltage. When used in this
mode, input to input isolation is not maintained.
7
When used with live contacts, each input shall be able to accept up to 30 volts DC. Zero volts to 0.8 volts are recognized as a
logic zero and 2.5 volts to 20 volts are recognized as logic one.
8
The interface to the quadrature encoder shall consist of two inputs, 15 volts DC at 200 mA (described in a later section), and
common.
9
At least one detector interface module is required.
10
8.6 V nominal.
11
The power supply has the ability to control the power to the scintillation detector:
when commanded by the CPU module,
when the processor on the module detects a condition that could harm the scintillation detector,
when the watchdog timer generates a reset.