High Speed Calibration
PROCEDURE
(For weigh scales using tachometer or encoder for speed). If speed input is a signal from a line down
contact, see Speed Calibration using Line down Contact.
The belt must be running at the highest speed during the High Speed Calibration.
Using the Ronan local display or HART communicator find the Main Menu;
Scroll down until you reach and highlight Calibration.
Access Calibration by pressing the right arrow key.
Scroll down until you reach and highlight Cal Speed.
Access the Cal Speed by pressing the right arrow key.
Scroll down until you reach and highlight High Calibrate.
Access the High Calibrate by pressing the right arrow key.
Access the High Speed Cal by pressing the right arrow key.
Press the F4 key (OK) to the prompt ‘Warning – Remove loop from automatic control
before doing the calibrate procedure’.
Wait until the prompt changes from
Performing Calibrate procedure’ … to ‘High speed calibrate SUCCESS. Remember to enter
speed’.
Press the F4 key (OK).
Scroll down until you reach and highlight High Speed 50.000 ft/min. (Note 50.000 ft/min value and units
may very depending on the what value and units were selected.) (Note: You must access the High Speed
and either change or acknowledge the value to complete the calibration.)
Access High Speed 50.0000 ft/min by pressing the right arrow key.
Use the arrow keys to select and change the display value to the correct speed. Once the displayed value is
correct, press the F4 key (Enter).
The Speed Calibration is complete.