Instructions
and
Operating Manual
X96S
DENSITY GAGE
60%
Ronan Engineering
Process Setup
Filtering
Low Alarm
High Alarm
Empty Clamp
Linearization
Temp.
Press.
Vue de la page 0
1 2 3 4 5 6 ... 66 67

Résumé du contenu

Page 1 - DENSITY GAGE

InstructionsandOperating ManualX96SDENSITY GAGE60%Ronan EngineeringProcess SetupFilteringLow AlarmHigh AlarmEmpty ClampLinearizationTemp.Press.

Page 2 - Table of Contents

Density Gage Manual 0124076Menus/OperationMenu TreesThe Ronan X96S Density Gage uses a tree structured menu system.Ronan X96S - DensityVariables Varia

Page 3 - Density Gage Manual 012407

Density Gage Manual 0124077ConfigurationOperation OperationD1 Config 3D2 Config 3 Filtering FilteringProc Temp Config Empty ClampPressure Config

Page 4

Density Gage Manual 0124078Configuration (Cont’d) Density 1 Config % Solids VariablesD1 Config D1 Units [units] Carrier [value]D2 Config Low Range [va

Page 5 - Overview

Density Gage Manual 0124079Configuration (Cont’d) Hardware (Cont’d)Hardware (Cont’d) System HardwareProc Temp [variable]Source TypePressureAnalog Out

Page 6 - Basic Concepts

Density Gage Manual 01240710Digital Outputs OutputOutput [Relay 1-4/TTL 1-4] Relay 1Select Sources [variable] Relay 2Polarity [variable] Relay 3Relay

Page 7 - Principles of Operation

Density Gage Manual 01240711CalibrationState [state]Ref Constants Ref ConstantsCalibrateLast Ref Date [date] Ref Mode [variable]Last Ref Time [time] C

Page 8

Density Gage Manual 01240712Root MenuThe root menu is titled “Ronan X96S – Density”. It contains the following items:ITEM FUNCTIONVariablesSelecting t

Page 9

Density Gage Manual 01240713Status Display MenuThe menu titled “Status Display” is used to configure the device status display. It contains the follow

Page 10 - Menus/Operation

Density Gage Manual 01240714Operation MenuThe Operation menu is used to access the menus and variables that control the processing of the density data

Page 11 - [enable/disable]

Density Gage Manual 01240715Empty Clamp MenuThe X96S uses a mechanism called empty clamp to protect detectors (particular scintillation detectors) fro

Page 12

Density Gage Manual 012407iTable of ContentsOVERVIEW...

Page 13

Density Gage Manual 01240716ITEM FUNCTIONTable Entry #Shows and allows the user to select an entry in the linearization tableEntry UsedShows if the en

Page 14 - [Relay 1-4/TTL 1-4] Relay 1

Density Gage Manual 01240717ITEM FUNCTIONPres CompShows the current state of the pressure compensation mechanism (enabled or disabled)and allows the u

Page 15

Density Gage Manual 01240718% Solids Variables MenuThe % Solids Variables menu is used to configure the parameters associated with the percent of soli

Page 16 - Root Menu

Density Gage Manual 01240719ITEMS MEANINGdegBalldegrees ballingPercent_StmQualpercent steam qualityPercent Acidspercent acids% Solids Variables MenuT

Page 17 - Configuration Menu

Density Gage Manual 01240720Pressure Config MenuThe Pressure Config menu is used to configure the parameters associated with the process pressure meas

Page 18 - Filtering Menu

Density Gage Manual 01240721Hardware MenuThe Hardware menu is used to define the type of hardware used to provide measurements and radiation. It conta

Page 19 - Linearization Menu

Density Gage Manual 01240722Source Type MenuThe Source Type menu is used to define the type of radiation source used. It contains the following items:

Page 20 - Pressure Comp Menu

Density Gage Manual 01240723HART Output MenuThe HART Output menu is used to define the type of HART Output desired.ITEM FUNCTIONX96S-2005Selecting thi

Page 21 - D1 Config Menu

Density Gage Manual 01240724Digital Outputs MenuThis menu is used to view and configure the digital outputs. It contains the following items:ITEM FUN

Page 22 - D2 Config Menu

Density Gage Manual 01240725Relay and TTL MenusThe Relay and TTL menus are used to configure the X96S Relay Outputs and the 4 TTL Outputs. The Relay a

Page 23 - Proc Temp Config Menu

Density Gage Manual 012407iiCom1 Protocol Menu...

Page 24 - Head Temp Config Menu

Density Gage Manual 01240726Calibration MenuThis menu is used to view and control the calibration of the X96S Density Gage. It contains the following

Page 25 - Hardware Menu

Density Gage Manual 01240727Loop Config MenuThis menu is used to access the primary 4-20ma loop calibration procedures. It contains the following ite

Page 26 - Analog Out Config

Density Gage Manual 01240728Calibrate MenuThis menu is used to access the various calibration procedures. It contains the following items:ITEM FUNCTI

Page 27 - System Menu

Density Gage Manual 01240729Pressure Reference MenuThis menu is used to perform the process pressure reference procedure. You will be asked to provid

Page 28 - Digital Outputs Menu

Density Gage Manual 01240730X96S Local DisplayNavigating MenusThe menu and the display screen are one or more lines, each consisting of a line label (

Page 29 - Digital Inputs Menu

Density Gage Manual 01240731Editing ValuesThe editing of different types of values is designed around the use of the four direction keys and up to 4 f

Page 30 - Calibration Menu

Density Gage Manual 01240732InstallationCautionSpecific License(SA or GS Series)General License(RLL Source Holder)UnpackingStorageRonan's Mold L

Page 31 - Aux Loop Cfg Menu

Density Gage Manual 01240733Inspection The source holder is equipped with an ON/OFF mechanism.During shipment and storage themechanism MUST BE SECURE

Page 32 - Calibrate Menu

Density Gage Manual 01240734Safety PrecautionsDuring installation the RSO will provide guidelines toassure safety. Consider the information presented

Page 33 - Diagnostics

Density Gage Manual 01240735Mechanical MountingReview the Configuration Drawing, which is included in theDrawing Chapter of this manual.Please referen

Page 34 - X96S Local Display

Density Gage Manual 012407iiiReference Modes 45Types of Calibration 45Calibration ...

Page 35 - Editing Values

Density Gage Manual 01240736Electrical Installation of Interconnect WiringDO NOT APPLY POWER until wiring is carefully checked.Wire the equipment acco

Page 36 - Installation

Density Gage Manual 01240737Microprocessor VerificationRotate latch clockwise to open the enclosure door. Nextremove the computer front cover by slid

Page 37 - ------- WARNING

Density Gage Manual 01240738Power-upCF1 F2F3F4 X96SBefore applying power, ensure all boards are fully seatedin frame’s slots. Close front door of the

Page 38 - Safety Precautions

Density Gage Manual 01240739PasswordNotice:To access the Programming Menu, the Password is 101010.Step 1: Power Up – You should now be on the Status S

Page 39 - Mechanical Mounting

Density Gage Manual 01240740Quick Start Reference – Calibrating DensityWhat type of calibration is needed?Two types of calibration are available for R

Page 40 - NATIONAL CODE

Density Gage Manual 01240741Quick Start Reference – Calibrating DensityStep 1 Start at the Status DisplayFrom the Status Display Screen, Press the F3

Page 41 - Microprocessor Verification

Density Gage Manual 01240742Quick Start Reference – Calibrating Density(For Completion of a Single Point Reference)CF1 F2 F3 F4Ronan EngineeringX96S D

Page 42 - Power-up

Density Gage Manual 01240743Quick Start Reference – Calibrating Density(For adjusting the 1/uT or Transmission Constant)The quick start calibration is

Page 43 - Password

Density Gage Manual 01240744Quick Start Reference – Calibrating Density(For Completing a Dual-Point Calibration)CF1 F2 F3 F4Ronan EngineeringX96S Dens

Page 44 - – Calibrating Density

Density Gage Manual 01240745CalibrationCalibration correlates the X96S's output to your actual process density. It instructs the microprocessor

Page 45 - Step 2 Password Menu

Density Gage Manual 0124071OverviewThe X96S is a family of measurement products that is intended to replace the current X96N and X99 product families.

Page 46 - F1 F2 F3 F4

Density Gage Manual 01240746Calibration ConstantThe two values, the reference density (do) and the calibration constant (1/ut), are used by the X96N a

Page 47 - Ronan Engineering

Density Gage Manual 01240747High Calibration (Dual Point Calibration)An alternate calibration technique is the "Dual Point Calibration" wher

Page 48

Density Gage Manual 01240748Single-Point Calibration ProcedureAs an alternative, the single-point calibration can be used. In that case, the results

Page 49 - Calibration

Density Gage Manual 01240749DocumentationFor future reference, document these items:(a) Environmental/process conditions (densities, temperatures,

Page 50 - Calibration Curve

Density Gage Manual 01240750ConfigurationRonan ships the Density Monitor System with factory-default software settings. Those settings are responsibl

Page 51 - Preparation for Calibration

Density Gage Manual 01240751DetectorScintillator DetectorDescription The Ronan scintillation detector consists of three maincomponents: The plastic sc

Page 52 - Temperature Compensation

Density Gage Manual 01240752Detector Service the critical components of the electronic circuit and thePMT/Crystal Assembly are aligned before leaving

Page 53 - Documentation

Density Gage Manual 01240753ION ChamberDetector/Amplifier Ronan’s ion chamber detector is filled with an inertAssembly high-pressure gas. It uses low

Page 54 - Configuration

Density Gage Manual 01240754Servicing the Detector The ion-chamber detector contains pressurized inertgas. The ion chamber itself is not serviceable

Page 55 - Detector

Density Gage Manual 01240755Detector Removal/ReplacementELONGATED DETECTORS NOTES:To avoid damage in shipment or installation, theelongated detectors

Page 56

Density Gage Manual 0124072Basic ConceptsCommunicationsThe Ronan X96S Density gage provides both 4-20 mA current loop and HART communications.4-20 MAF

Page 57 - ION Chamber

Density Gage Manual 01240756Removing the DetectorAmplifier Circuit Board(CBAY-6102)Follow this procedure to remove the electrometer amplifier circuitb

Page 58

Density Gage Manual 01240757Replacing the DetectorCircuit Board/ConnectorAssemblyReference: B-6102-KIf installing a new electrometer amplifier

Page 59

Density Gage Manual 01240758ElectronicsX96-2001PLX96-2001PL is the CPU moduleX96-2003-01PLX96-2003-01PL is the Ionization Chamber InputX96-2003-02PLX9

Page 60 - (CBAY-6102)

Density Gage Manual 012407594 relay (2 Amp capacity) output points are provided. Form “C” outputs are brought out to the connector (three connections

Page 61 - Assembly

Density Gage Manual 01240760OptionsX96S Mechanical Chassis & LCD Part NumbersPART NUMBER DESCRIPTIONCHAS-0511-6 X96S-N4-1, NEMA 4 Enclosure, 6 Pos

Page 62 - Electronics

Density Gage Manual 01240761SPECIFICATIONSMODEL X96SProcess Computer: Microprocessor-based unit wit a liquid crystal display, push-button interface,HA

Page 63

Density Gage Manual 01240762F1CF2 F3 F4 X96SSERIES X96S-N4

Page 64 - PART NUMBER DESCRIPTION

Density Gage Manual 01240763RONAN ENGINEERING COMPANY21200 Oxnard StreetWoodland Hills, California 91367 U.S.A.(800) 327-6626 • FAX (818) 992-6435E-Ma

Page 65 - SPECIFICATIONS

Density Gage Manual 0124073TheoryTheory of Radiation GagingRadiation gages operate on the principle of radiation absorption and transmission.A beam of

Page 66 - SERIES X96S-N4

Density Gage Manual 0124074• Digital Filtering – signal smoothing to reduce statistical radiation noise• Dynamic Tracking – quick gage response to q

Page 67

Density Gage Manual 0124075PasswordNotice:To access the Programming Menu, the Password is 101010.Step 1: Power Up – You should now be on the Status Sc