15
ITEM FUNCTION
Table Entry #
Shows, and allows the user to select, an entry in the linearization table
Entry Used
Shows if the entry is used or not.
Measured
Shows, and allows the user to set, the measured value associated with this linearization
table entry. This is a value calculated by the X96S.
Actual
Shows, and allows the user to set, the actual value associated with this linearization
table entry. This value is the result of actual level knowledge, and compares to the
Measured value above.
Set Entry
This item invokes a method that sets a table entry
Remove Entry
This item invokes a method that removes a table entry
Level Config Menu
The Level Config menu is used to configure the parameters associated with the level measurement. It contains the following
items:
ITEM FUNCTION
Units
Shows, and allows the user to set, the level units used
Low Range
Shows, and allows the user to set, the level value to be mapped to 4ma on the current
loop output, if level is selected to control that current loop.
High Range
Shows, and allows the user to set, the level value to be mapped to 20ma on the current
loop output, if level is selected to control that current loop.
Units is one of the following:
Units MEANING
ft
feet
m
meter
in
inch
cm
centimeter
mm
millimeter
Head Temp Config Menu
The Head Temp Config menu is used to configure the parameters associated with the detector electronics temperature
measurement. This function is used primarily in high-temperature applications where the temperature exceeds the electronics
temperature specifications. It contains the following items:
ITEM FUNCTION
Temp Units
Shows, and allows the user to set, the units to be used for head temperature
Low Range
Shows, and allows the user to set, the temperature value to be mapped to 4ma on the
current loop output, if head temperature is selected to control that current loop.
High Range
Shows, and allows the user to set, the temperature value to be mapped to20ma on the
current loop output, if head temperature is selected to control that current loop.