Page 1 of 2
Rev. 0
A Family of
Products!!!
Transmitters
Alarm Trips
I/I Isolators
Ronan Engineering Company
Ph: 800-327-6626 Fax: 818-992-6435 E-mail: sales@ronan.com Web: www.ronan.com
High Density Four (4) Channel Packaging
Input Loop Powered
Low Input Loop Burden (<6 vdc @ 250 ohm load)
Ideal for High Density I-to-I Isolator Requirements
Max. Output Loop Load = 500 ohms
Full Input/Output/Power and Channel Isolation
Standard 35 mm DIN Rail Mounting
DESCRIPTION: The X54-3224 is a four channel current-to-
current repeater/isolator transmitter powered by the input 4-20 mA cur-
rent loops. Features include: two piece plug-in connectors for easy wir-
ing and maintenance; full 1000 vac isolation; and front panel access to
ZERO and SPAN Adjustments. Dual output options per channel al-
lows for matching load requirements and Input loop burden. RFI pro-
tection, wide operating temperature, and high accuracy are standard
features.
4-20 mA Signal Repeater/Isolator Transmitters - DIN Rail Mounting
X54-3224 Four (4) Channel Input Loop Powered I to I Isolator (mA transmitter)
ISO 9001
REGISTERED
SPECIFICATIONS:
Inputs:
4-20 mA (input loop burden is <6 vdc drop+(IloadXRload))
Outputs:
Output #1 = 4-20 mA into 0 to 99 ohm load (3 vdc)
Output #2 = 4-20 mA into 100 to 500 ohm load (10 vdc)
Span Adjustment:
Front-accessible, multi-turn, infinite resolution potentiometer
permits minimum +/- 5% adjustment.
Zero Adjustment:
Front-accessible, multi-turn, infinite resolution potentiometer
permits minimum +/- 5% adjustment
Calibrated Accuracy: +/- 0.1% of span
Isolation: 1,000 vac between channels, input and output.
Common Mode Rejection: > 120 dB, DC at 60 Hz.
Common Mode Voltage:
1000 volts peak AC maximum without damage.
Ambient Temperature Coefficient:
Ambient temperature range: 32 to 158 F (0 to 70 C).
Gain: < +/- 0.01 %/ F.
Zero: < +/- 0.005 %/ F.
Ambient Temperature Range:
Operating: -20 to 158 F (-25 to 70 C).
Storage: -40 to 158 F ( -40 to 70 C).
Radio Frequency Effects:
< 0.4 mV (referred to input) + 0.2% of span (referred to out-
put) when exposed to 5W transmitter with frequency range
20-460 MHz at a distance of meter.
Terminals:
High temperature polyester type, wire size 12 AWG max., 10
A max., 300 V max.
Specifications apply at 23 +/- 2°C (74 +/- 2°F) unless otherwise specified, and are subject to change without notice,