Ronan X96S DENSITY GAGE Manuel d'utilisateur Page 54

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 67
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 53
Density Gage Manual 012407
50
Configuration
Ronan ships the Density Monitor System with factory-default software settings. Those settings are responsible for the information
that initially appears on the status displays.
After installation at your site, you may need to reconfigure the system to fit your application. The goal is to correlate the X96S
output with your lab sampling, so the X96S density monitor can replace the time-consuming task of constant manual lab sampling.
The list below summarizes the activities that are detailed in the remainder of this chapter:
- Check the factory-default settings to be sure they are appropriate for your circumstances. IF NOT, make the
necessary changes and document those changes for future reference.
- Enable and setup temperature compensation before doing the calibration, if required.
- Perform an initial calibration to correlate the X96S’s output to the actual process density as determined by
laboratory sampling.
- Document detector output counts at calibrated values to assist in troubleshooting. Also, record changes you
make to factory-default settings. Keep this information for future reference.
Vue de la page 53
1 2 ... 49 50 51 52 53 54 55 56 57 58 59 ... 66 67

Commentaires sur ces manuels

Pas de commentaire